Andreas Schmid Climate Chamber -40 °C to +130 °C

The test chamber is a monoblock construction.
It consists of the actual test space with the recirculating air mixing chamber and the necessary heat exchanger, as well as the unit frame, which is arranged behind the test space and is a fixed component of the system.
Control and regulation are located on the front side of the unit frame.

Chamber design: Monoblock, welded

Dimensions

  • Test space: (W x D x H) 3000 x 1150 x 1600 mm
  • External dimensions: (W x D x H) 3740 x 1390 x 1880 mm
  • Unit frame: (W x D x H) 1500 x 1000 x 1000 mm
  • Door double leaf: (W x H) 3000 x 1600 mm
  • Test space window, heated: (W x H) 300 x 500 mm

Interior lining

  • Walls and ceiling: V2A sheet metal, 1 mm smooth
  • Floor: V2A sheet metal, 2 mm embossed
  • Floor load: 20000 N/m²
  • Wheel load: 3000 N/4cm², 200 mm rubber wheel
  • Insulation: 120 mm PUR, CFC-free B1 according to DIN 4102
  • Exterior cladding: steel sheet 0.6 mm, painted and to RAL 9002
  • Temperature range: -40 °C to +130 °C
  • Temperature constancy +- 1.0 K - temporal
  • Climate range: +10 °C to +90 °C
  • Humidity constancy: +-3% r.h.

Average rate of temperature change: 3K/min on average
from: -40 °C to +130 °C and vice versa

Humidifier: Steam humidifier for operation with fully demineralized water supplied on-site

Refrigeration supply: refrigeration compressor, water-cooled
Refrigerant: R 404 A
Cooling water inlet temperature: 15 °C
Ambient temperature: 35 °C
Heat supply: electric resistance heating

Electrical data
Voltage: 400 V / 3 Ph / N + Pe / 50 Hz

The current consumption depends on the operating conditions for the:

Cooling operation: 31 A
Heating operation: 26 A
Climate operation: 50 A


Further items - new and used - can be found in our shop! International shipping costs on request!


Technische Daten

Chamber design: Monoblock, welded

Dimensions
Test space: (W x D x H) 3000 x 1150 x 1600 mm
External dimensions: (W x D x H) 3740 x 1390 x 1880 mm
Unit frame: (W x D x H) 1500 x 1000 x 1000 mm
Door double leaf: (W x H) 3000 x 1600 mm
Test space window, heated: (W x H) 300 x 500 mm

Interior lining
Walls and ceiling: V2A sheet metal, 1 mm smooth
Floor: V2A sheet metal, 2 mm embossed
Floor load: 20000 N/m²
Wheel load: 3000 N/4cm², 200 mm rubber wheel
Insulation: 120 mm PUR, CFC-free B1 according to DIN 4102
Exterior cladding: steel sheet 0.6 mm, painted and to RAL 9002

Temperature range: -40 °C to +130 °C
Temperature constancy +- 1.0 K - temporal
Climate range: +10 °C to +90 °C
Humidity constancy: +-3% r.h.

Average rate of temperature change: 3K/min on average
from: -40 °C to +130 °C and vice versa

Humidifier: Steam humidifier for operation with fully demineralized water supplied on-site

Refrigeration supply: refrigeration compressor, water-cooled
Refrigerant: R 404 A
Cooling water inlet temperature: 15 °C
Ambient temperature: 35 °C
Heat supply: electric resistance heating

Electrical data
Voltage: 400 V / 3 Ph / N + Pe / 50 Hz

The current consumption depends on the operating conditions for the:

Cooling operation: 31 A
Heating operation: 26 A
Climate operation: 50 A

Rechtliche Hinweise – Klicken Sie hier für weitere Informationen

Ihre Privatsphäre ist uns wichtig
Wir verwenden Cookies und ähnliche Technologien auf unserer Website und verarbeiten personenbezogene Daten von Besucher:innen unserer Webseite (z.B. IP-Adresse), um z.B. Inhalte zu personalisieren, Medien von Drittanbietern einzubinden oder Zugriffe auf unsere Website zu analysieren. Die Datenverarbeitung erfolgt erst durch gesetzte Cookies. Wir teilen diese Daten mit Dritten, die wir in den Einstellungen benennen. Die Datenverarbeitung kann mit Einwilligung oder aufgrund eines berechtigten Interesses erfolgen. Die Zustimmung kann erteilt oder abgelehnt werden. Es besteht das Recht, nicht einzuwilligen und die Einwilligung zu einem späteren Zeitpunkt zu ändern oder zu widerrufen. Beachten Sie unser Impressum und weitere Hinweise zur Verwendung personenbezogener Daten in unserer Datenschutzerklärung.